AtriClip FLEX Device

Request More Information

Product Features

  • Flexible 6 cm shaft
  • Closed-ended clip
  • Plunger grip
  • Available in four sizes

The AtriClip LAA Exclusion System is indicated for the exclusion of the heart's left atrial appendage, performed under direct visualization an in conjunction with other cardiac surgical procedures.

INTERESTED IN LEARNING MORE?

 
PRODUCT CODES:
  • ACH235 (35 mm)
  • ACH240 (40 mm)
  • ACH245 (45 mm)
  • ACH250 (50 mm)
Page References
MRI SAFETY INFORMATION:

MR CONDITIONAL

The LAA Exclusion System Clip is MR Conditional. A patient with the LAA Exclusion System Clip may be safely scanned under the following conditions. Failure to follow these conditions may result in injury to the patient:

  • Nominal Values of Static Magnetic Field: 1.5-Tesla or 3.0-Tesla
  • Maximum Spatial Field Gradient: 40 T/m (4,000 gauss/cm)
  • Type of RF Excitation: Circularly Polarized (CP) (i.e., Quadrature Transmission)
  • Transmit RF Coil Information: There are no transmit RF coil restrictions.
  • Operating Mode of MR System: Normal Operating Mode
  • Maximum Whole Body Averaged SAR: 2-W/kg (Normal Operating Mode)
  • Limits on Scan Duration: Whole body averaged SAR of 2-W/kg for 60 minutes of continuous RF exposure (i.e., per pulse sequence or back-to-back sequences/series without breaks)
  • MR Image Artifact: The presence of this implant produces an imaging artifact. Therefore, carefully select pulse sequence parameters if the implant is located in the area of interest.
PM-US-0306E-0826-G